Transdermal drug delivery TDD systems have many advantages but are conventionally limited by the low permeability of skin. The idea of using microneedles to painlessly penetrate the topmost impermeable stratum corneum has previously been put forward. In this paper, the fabrication of solid and hollow silicon microneedles with straight side-walls and with the following dimensions: 20-100 microm in diameter and 100-150 microm in length is described. In vitro tests demonstrate that with prior solid microneedle application, transdermal drug transport is significantly increased by 10-20 times, with the degree of enhancement being related to needle diameter. In vivo tests in diabetic animals, however, were unable to demonstrate any delivery of insulin through the hollow microneedles. It is proposed that two factors, microneedle length and tip sharpness, have to be improved for systemic drug delivery to be seen in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.