Abstract

Damaged optics of an intraocular lens (IOL) can impair visual functioning. Damage can occur during the manufacturing process, during implantation, or during post-implantation Nd:YAG laser posterior capsulotomy. An in vitro experiment was performed to identify the potential IOL damage patterns associated with the use of the Nd:YAG laser for posterior capsulotomy. A review of the literature regarding Nd:YAG laser damage to IOLs was performed. The objectives were to describe the damage patterns to a sample IOL by performing virtual posterior capsulotomies. Patterns of damage were demonstrated and described. Although optic damage can occur with use of the Nd:YAG laser, it is infrequent and usually the result of improper depth of focus.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.