Abstract

SELDI-TOF MS is a mass spectrometric technique which has been extensively used for biomarker discovery. In this study, we show that in-source decay is an important source for the generation of additional spectral peaks with this technique, both for pure proteins and proteins in serum samples. Thus, SELDI-TOF MS could be used to gain sequence information from proteins, but the results also question the uncritical use of SELDI-TOF MS as a general method for the detection of biomarkers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.