Abstract

We demonstrate the feasibility of resetting and reusing dosimeters exploiting the measurement of the infrared radiation-induced attenuation (IR-RIA) in phosphosilicate optical fibers (OFs) to provide point or distributed dose measurements in radiation environments. To bleach the room temperature stable IR-RIA, we used the photobleaching (PB) phenomenon. The PB efficiency was evaluated for different wavelengths in the [400-1100] nm range. The best identified PB resetting condition consists in using a continuous-wave Argon-ion laser at 514 nm. This treatment successfully bleached ∼97% of the IR-RIA at 1550 nm of a 30 m-long P-doped single mode optical fiber X-ray irradiated at a dose of 100 Gy. Successive re-irradiations of the same OF sample, regenerated after each run, confirm that the dosimeter keeps the same calibration curve during the whole process.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.