Abstract

Measurement of nuclear materials by ISOCS technique an important and significant tool for IAEA activities verification. ISOCS technique able to measuring samples with different geometry. In this work, ISOCS technique was used to calculate the absolute efficiency of the Hyper Pure Germanium (HPGe) detector. The calculated absolute efficiency of the detector was used in combination with experimental work to determine U-235 mass in cylindrical NM samples. A good agreement was obtained when compared estimated and declared values of 235U mass content with difference in the ranged from −0.827% to 1.635%.