Abstract

We report on the development of a submersible probe for the simultaneous potentiometric detection of carbonate, calcium, and pH in seawater. All-solid-state electrodes incorporating nanomaterials provide an adequate response time (<10 s), stability (drifts of <0.9 mV h–1), reproducibility (calibration parameter deviation of <0.7%), and accuracy (deviation of <8% compared to reference techniques) for real-time monitoring of seawater using a flow system. The functioning of the deployable prototype was checked in an outdoor mesocosm and via long-term monitoring in Genoa Harbor. The electrodes worked properly for 3 weeks, and the system demonstrated the capability to autonomously operate with routines for repetitive measurements, data storage, and management. In situ profiles observed in Genoa Harbor and Arcachon Bay were validated using on site and ex situ techniques. The validation of in situ-detected carbonate is a challenge because both re-equilibration of the sample with atmospheric CO2 and the use of ap...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.