A method has been proposed for the determination of mercury by cold vapor generation graphite furnace atomic absorption spectrometry (CV-GFAAS) with Pd-Rh as coating and chemical modifier. The trapping efficiency for mercury with Pd-Rh was higher than with Pd alone. The characteristic mass of the method, which gives an integrated absorbance of 0.0044 s, was found to be 55 pg and the absolute detection limit (3 σ) of 37 pg was obtained with the proposed modifier. The method was successfully applied to the determination of mercury in standard reference water samples, wastewater samples and cosmetics with a recovery range of 95–104%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.