Abstract

A 600W helium refrigerator/liquefier with variable temperature supplies was constructed in the National Institute for Fusion Science (NIFS). The mass flow rate of supply Supercritical Helium (SHe) is important information for its performance tests and experimental uses. A quarter circle orifice, according to Verein Deutscher Ingenieure and Verband Deutscher Elektrotechniker (VDI/VDE) 2041, is equipped as a mass flow meter. In the case of the configuration of our system, it is recommended to calibrate the orifice under identical operation conditions with regard to Reynolds number and installation. Therefore, a new in-situ calibration method of the orifice is developed. A dummy heat load was attached between SHe supply and return ports, and the resulting SHe enthalpy rise gives us the information for a mass flow rate. The obtained coefficients of discharge are consistent with those of previous studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.