Abstract

To investigate the chemical changes occurring at metal/lubricant interfaces under tribological conditions in the boundary-lubrication regime, an in situ system for conducting quantitative tribological measurements has been constructed by combining an attenuated total reflection Fourier-transform infrared (ATR/FT-IR) spectrometer with a reciprocating tribometer. By periodically acquiring ATR/FT-IR spectra during sliding, spectroscopic changes due to thermal and/or tribochemical reactions occurring at the metal/oil interface can be monitored and correlated with friction measurements. The usefulness of this tribological test system has been demonstrated by performing ATR tribological experiments in the presence of a poly-α-olefin base oil at high temperature (423 K) on iron-coated germanium ATR crystals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.