Abstract

In the clinical application of intravascular optical coherence tomography (IVOCT), it is necessary to flush opaque blood during image acquisition. However, there are no specific standards for how to perform low-dose but effective flushing. In this study, computational fluid dynamics (CFD) and optical models were integrated to numerically simulate the complete process of IVOCT, which includes blood flushing with normal saline followed by image acquisition. Moreover, an intermittent injection scheme was proposed, and its advantages over the conventionally adopted scheme of continuous injection were verified. The results show that intermittent injection can significantly reduce the dosage of normal saline (reduced by 44.4%) with only a slight sacrifice of image quality (reduced by 8.7%, but still acceptable). The developed model and key findings in this work can help surgeons practice optimized IVOCT operations and potentially lead to improved designs of the IVOCT equipment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.