Abstract

Non-invasive cardiac output methods such as Electrical Cardiometry (EC) are relatively novel assessment tools for neonates and they enable continuous monitoring of stroke volume (SV). An in-silico comparison of differences in EC-derived SV in relation to preset length and weight was performed. EC (ICON, Osypka Medical) was simulated using the “demo” mode for various combinations of length and weight representative of term and preterm infants. One-centimetre length error resulted in a SV-change of 1.8–3.6% (preterm) or 1.6–2.0% (term) throughout the tested weight ranges. One-hundred gram error in weight measurement resulted in a SV-change of 5.0–7.1% (preterm) or 1.5–1.8% (term) throughout the tested length ranges. Algorithms to calculate EC-derived SV incorporate anthropomorphic measurements. Therefore, inaccuracy in physical measurement can impact absolute EC measurements. This should be considered in the interpretation of previous findings and the design of future clinical studies of EC-derived cardiac parameters in neonates, particularly in the preterm cohorts where a proportional change was noted to be greatest.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.