Abstract
In this study, we present a sensorless, robust, and physiological tracking control method to drive the operational speed of implantable rotary blood pumps (IRBPs) for patients with heart failure (HF). The method used sensorless measurements of the pump flow to track the desired reference flow (Qr). A dynamical estimator model was used to estimate the average pump flow (Q^est) based on pulse-width modulation (PWM) signals. A proportional-integral (PI) controller integrated with a fuzzy logic control (FLC) system was developed to automatically adapt the pump flow. The Qr was modeled as a constant and trigonometric function using an elastance function (E(t)) to achieve a variation in the metabolic demand. The proposed method was evaluated in silico using a lumped parameter model of the cardiovascular system (CVS) under rest and exercise scenarios. The findings demonstrated that the proposed control system efficiently updated the pump speed of the IRBP to avoid suction or overperfusion. In all scenarios, the numerical results for the left atrium pressure (Pla), aortic pressure (Pao), and left ventricle pressure (Plv) were clinically accepted. The Q^est accurately tracked the Qr within an error of 0.25 L/min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.