Abstract. The Plasma Electron and Current Experiment (PEACE) instruments operate on all four of the Cluster spacecraft and measure the 3-D velocity distribution of electrons in the energy range from 0.59 eV to 26.4 keV during each spacecraft spin. Pitch angle distributions and moments of the velocity distribution are also produced. As the mission has progressed, the efficiency of the detectors has declined. Several factors may play a role in this decline such as exposure to radiation, high electron fluxes and spacecraft thruster firings. To account for these variations, continuous in-flight calibration work is essential. The purpose of this paper is to describe the PEACE calibration parameters, focussing in particular on those that vary over time, and to describe the methods which are used to determine their evolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Geoscientific Instrumentation, Methods and Data Systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.