Abstract

Routine quality control procedures for mammography need a periodical testing of the X ray units. The checks of kVp accuracy, mA linearity, exposure reproducibility and HVL consistency require several instruments (non-invasive kVp meter, dosemeter, absorbers, etc.) and a lot of time. The use of a new instrument for a quick and easy measurement of kVp ripple, HVL, exposure time and incident air kerma is presented. The instrument, composed of an X ray probe and a portable computer, allows the evaluation of the above parameters in a single shot. Two mammographic units have been routinely tested measuring kVp, HVL, exposure time and incident air kerma. The evaluation has been performed exposing the probe during the execution of a performance phantom radiograph. The relationships between the fluctuations of the measured parameters are compared with the limiting values proposed by CEC guidelines. Correlation analysis of the whole data set allows malfunctioning identification. The instrument works properly and is very useful in quality control activity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.