Abstract
Abstract. The in‐field calibration of a dielectric probe to measure soil water content is described. The probe uses an access tube analogous to that of the neutron probe. The dielectric constant was measured at soil depths of 10, 20, 30, 40, 60 and 100 cm. Cores of soil were then taken from the face of pits dug 30 cm from the access tube and their soil water contents determined by oven drying. The dielectric constant values measured by the probe were calibrated against water contents from these cores. We found that sensor depth needed to be included to achieve a good calibration model that explained 72% of the variance. It is argued that depth needs to be included because of artefacts introduced during the installation of the access tube.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.