Abstract

Spectrophotometric pH measurements have been widely used in oceanic CO 2−system research. However, impurities in indicator dyes from different manufacturers may cause uncertainty in measured pH values. In this work we report initial findings surrounding the potential significance of indicator sources (manufacturers) on pH measurements obtained with the sulfonephthalein indicator m-cresol purple. HPLC analyses were used for comparative assessments of impurities in m-cresol purple obtained from a range of manufacturers. Results indicate that m-cresol purple from different manufacturers, and perhaps from different batches of the same manufacturer, have different types and quantities of light-absorbing impurities. Impurities can contribute to pH offsets as large as 0.01 pH units. The pH offset caused by impurities decreases with decreasing sample pH. Indicators can be purified using preparative HPLC. Until purified indicators and refined indicator calibrations become available, we suggest that investigators should reserve some portion of each indicator batch for comparison with the properties of purified indicators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.