Abstract

The influence of the law of distribution of the measurand on the length parameter for a guard band when assessing compliance is considered. The existing algorithm for expanded measurement uncertainty evaluation described in the current regulatory documents and its shortcomings is analyzed. The features of uncertainty evaluation in calibrations and the possibility of using the kurtosis method for its implementation are investigated. The proposed approach is applied to the examples described in Supplement 2 to the EA-4/02 M:2013.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.