In the presented work, the leading edge timing and constant fraction discrimination timing techniques were applied to beta rays emitted from a 137Cs radioactive source through the developed spectrometers in order to improve the beta energy spectrum of the source. The energy spectra, with and without the techniques, were obtained with a BC‐400 plastic scintillation detector. The obtained spectra were compared with each other for the elimination of spurious pulses, which rise from background, backscattering, scattering, and any radiations other than the source, on the multichannel analyzer output.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.