Abstract
Potentiometric sensors will generally suffer from unwanted responses as a result to changing temperatures by generating an electromotive force. Typically, this voltage drift has a non-linear character and therefore it is difficult to compensate using linear algorithms implemented in the analogue domain. A solution is proposed to improve the sensor characteristics by combining the digitized output of two CO2 rubidium silver iodide sensors with a specially designed digital algorithm to improve the limits of detection (LOD). Experiments show that this method has the capability to improve the LOD of the sensor with a factor 4.5x during temperature variations of 22 °C over a measurement period of 22 h. It enables potentiometric sensors to be used in low power wireless sensor networks for long term air quality control. Furthermore, the influence of depletion of the rubidium silver iodide electrolyte layer can be effectively compensated by determining the decay of the active layer according to the Nernst equation. Knowing the function of depletion over time helps to correct the sensor output and thereby improves the accuracy of the sensor.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.