Abstract

The colorimetric variations induced upon changes in interfacial refractive index of nanoscale noble metal structures exhibiting localized surface plasmon resonance (LSPR) provides a convenient means of label-free, affinity-based detection of biomolecular recognition reactions. However, despite being similar in nature to conventional SPR, LSPR has so far suffered from significantly lower data quality in terms of its signal-to-noise ratio (S/N) in typical biomolecular recognition analysis. In this work, generic data analysis algorithms and a simple experimental setup that provide a S/N upon protein binding that is comparable to that of state-of-the art SPR systems are presented. Specifically, it is demonstrated how temporal variations (rate approximately 0.5 Hz) in parameters proportional to the resonance peak position can be recorded simultaneously, yielding a peak position precision of <5 x 10(-4) nm and an extinction noise level of <5 x 10(-6) absorbance units (Abs). This, in turn, is shown to provide a S/N of approximately 2000 (equivalent to a detection limit of <0.1 ng/cm(2)) for typical protein binding reactions. Furthermore, the importance of utilizing changes in both peak position and magnitude is highlighted by comparing different LSPR active noble metal architectures that respond differently to bulk and interfacial refractive index changes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.