Abstract

Thermal ablation by ultrasound is being explored as a local therapy for cancers of soft tissue, such as the liver or breast. One challenge for these treatments are off-target effects, including heating outside of the intended region or skin burns. Improvements in heating efficiency can mitigate these undesired outcomes, and here, we describe methods for using phase-shift nanodroplets (PSNDs) with multi-focus sonications to enhance volumetric ablation and ablation efficiency at constant powers while increasing the pre-focal temperature by less than 6 C. Multi-focus ablation with 4 foci performed the best and achieved a mean ablation volume of 120.2 ± 22.4 mm3 and ablation efficiency of 0.04 mm3 J−1 versus an ablation volume of 61.2 ± 21.16 mm3 and ablation efficiency of 0.02 mm3 J−1 in single focus case. The combined use of PSNDs with multi-focal ultrasound presented here is a new approach to increasing ablation efficiency while reducing off-target effects and could be generally applied in various focused ultrasound thermal ablation treatments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.