We hypothesize that the combination of ultrasound Short-Lag Spatial Coherence (SLSC) imaging and Multi-Line Transmission (MLT) can provide enhanced visibility of highly-coherent targets within soft tissues, improving the frame-rate at the same time. Images of a biopsy needle inserted within a lab-made agar/gelatin phantom and an ex vivo bovine meat sample were acquired after implementing single-line transmission (SLT) and MLT with an increasing number of simultaneously transmitted beams. The received signals were then beamformed with standard Delay and Sum (DAS) and SLSC algorithms. Results show that MLT SLSC images have darker background than SLT SLSC and DAS B-mode images, which is due to the rapid signal decorrelation at short lags caused by inter-beam interferences. This effect in turn provides increased needle contrast with MLT SLSC imaging, particularly as the number of multiple beams increase, which is promising for a more accurate biopsy needle detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.