Abstract

PurposeTo use statistical process control for intensity-modulated radiation therapy (IMRT) quality assurance (QA) and improve tolerance limits and action limits. MethodsAn electronic portal imaging device (EPID) was selected to verify IMRT QA. The I-chart and the exponentially weighted moving averages (EWMA) chart were used to analyze the corresponding results. ResultsTwenty samples were used to enable the sampling requirements for building the control limits to be met. The I-chart showed that isolated data points beyond the control limits were mainly derived from complex plans. The EWMA made predictions of systematic errors earlier than the I-chart. Systematic errors primarily originated from the dose calibration on the EPID, and recalibrating the EPID could eliminate such errors. ConclusionStatistical process control is an effective tool to detect controllable and can be used in IMRT QA. After calibrating the EPID, the tolerance and action limits all improved and satisfied the requirements/recommended values of the AAPM TG-218 report.