Abstract
PurposeTo use statistical process control for intensity-modulated radiation therapy (IMRT) quality assurance (QA) and improve tolerance limits and action limits. MethodsAn electronic portal imaging device (EPID) was selected to verify IMRT QA. The I-chart and the exponentially weighted moving averages (EWMA) chart were used to analyze the corresponding results. ResultsTwenty samples were used to enable the sampling requirements for building the control limits to be met. The I-chart showed that isolated data points beyond the control limits were mainly derived from complex plans. The EWMA made predictions of systematic errors earlier than the I-chart. Systematic errors primarily originated from the dose calibration on the EPID, and recalibrating the EPID could eliminate such errors. ConclusionStatistical process control is an effective tool to detect controllable and can be used in IMRT QA. After calibrating the EPID, the tolerance and action limits all improved and satisfied the requirements/recommended values of the AAPM TG-218 report.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.