Abstract

While some MRI systems offer a "pause" function, combining it with the PROPELLER method for image quality improvement remains underexplored. This study investigated whether repositioning the head after pausing during PROPELLER imaging enhances image quality. All brain phantom images in this study were obtained using a 3.0T MRI and acquired using the fast spin-echo T2WI-based PROPELLER with motion correction. By combining the angle of rotational motion of the head phantom and the number of repositioning after a pause, two studies including seven trials were performed. Increasing the rotation angle decreased the image quality; however, pausing the image and repositioning the head phantom to the original angle improved the image quality. A similar result was obtained by repositioning the angle closer to its original angle. Experiments with multiple head movements showed that pausing the scan and repositioning the phantom with each movement improved image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.