Abstract

BackgroundTo improve accuracy of IOLMaster (Carl Zeiss, Jena, Germany) in corneal power measurement after myopic excimer corneal refractive surgery (MECRS) using multivariate polynomial analysis (MPA).MethodsOne eye of each of 403 patients (mean age 31.53 ± 8.47 years) was subjected to MECRS for a myopic defect, measured as spherical equivalent, ranging from − 9.50 to − 1 D (mean − 4.55 ± 2.20 D). Each patient underwent a complete eye examination and IOLMaster scan before surgery and at 1, 3 and 6 months follow up. Axial length (AL), flatter keratometry value (K1), steeper keratometry value (K2), mean keratometry value (KM) and anterior chamber depth measured from the corneal endothelium to the anterior surface of the lens (ACD) were used in a MPA to devise a method to improve accuracy of KM measurements.ResultsUsing AL, K1, K2 and ACD measured after surgery in polynomial degree 2 analysis, mean error of corneal power evaluation after MECRS was + 0.16 ± 0.19 D.ConclusionsMPA was found to be an effective tool in devising a method to improve precision in corneal power evaluation in eyes previously subjected to MECRS, according to our results.

Highlights

  • To improve accuracy of IOLMaster (Carl Zeiss, Jena, Germany) in corneal power measurement after myopic excimer corneal refractive surgery (MECRS) using multivariate polynomial analysis (MPA)

  • Axial length (AL), flatter keratometry value (K1), steeper keratometry value (K2), mean keratometry value (KM) and anterior chamber depth measured from the corneal endothelium to the anterior surface of the lens (ACD) were used in a MPA to devise a method to improve accuracy of KM measurements

  • MPA was found to be an effective tool in devising a method to improve precision in corneal power evaluation in eyes previously subjected to MECRS, according to our results

Read more

Summary

Introduction

To improve accuracy of IOLMaster (Carl Zeiss, Jena, Germany) in corneal power measurement after myopic excimer corneal refractive surgery (MECRS) using multivariate polynomial analysis (MPA). Conclusions: MPA was found to be an effective tool in devising a method to improve precision in corneal power evaluation in eyes previously subjected to MECRS, according to our results. A precise measurement of corneal power in patients who have previously undergone MECRS can help us to understand whether undercorrection or overcorrection is due to an error in the excimer laser calibration or in the evaluation of refraction before surgery, in order to design better ablation profiles for this type of surgery [16, 23]

Objectives
Methods
Results
Conclusion