In the standardization of 134Cs by the 4πβ−γ coincidence method with a γ-window at 605keV, the satellite components of 563keV and 569keV overlapping the 605keV peak cause a steep slope and non-linearity of the efficiency extrapolation function. By shifting the lower threshold of a γ-window higher, the satellite components are eliminated, and the slope tends to horizontal. Nearly flat efficiency curves were obtained by using a CeBr3 scintillator for detecting γ-photons, as well as a NaI(Tl) scintillator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.