Abstract
We have improved the ac-dc transfer differences and enhanced the sensitivity in our thin-film multijunction thermal converters (TFMJTCs) in vacuum. The ac-dc transfer differences of the TFMJTCs are less than 1 $\mu $ V/V down to 1 Hz and 10 $\mu $ V/V down to 0.2 Hz owing to the increased thermal time constant. The measured output voltage exceeds 120 mV at $10^{\mathrm {-4}}$ Pa owing to the reduction of heat conduction to the air. The sensitivity of the device, defined as the ratio of the output voltage to the input power, is 2.4 V/W, which is approximately three times larger than that measured at ambient pressure. Numerical simulations of the heat transfer of our TFMJTC explain the experimental results well.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.