Abstract
The effects of either static or pulsed magnetic fields on the reaction rate of Fremy's salt–ascorbic acid were studied directly by EPR spectroscopy. Radical pair mechanism (RPM) accounts for the magnetic field effects, but the expected amounts are so small that they need to be observed with particular care with EPR technique. The method is based on the resolution of a pair of EPR signals by the addition of a stationary field gradient, where the signals are coming from the exposed and control capillary sample. To this purpose, a suitable device for the gradient generation was used. Others improvements were the strictly keeping of the same boundary temperature condition in the capillary pairs, obtained by a refrigerating system controlled by a thermocouple, and the use of a pair of Helmholtz coils to generate an external high homogeneous magnetic field. By this experimental set up, we found that the magnetic field induce the decrease of the studied radical reaction rate. This EPR approach is a significant alternative to the spectrophotometric one. Moreover, it offers the advantage to detect both the radicals and/or intermediates involved in the reaction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.