Abstract
A gallium (Ga) focused ion beam (FIB) has been applied increasingly to 'site-specific' preparation of cross-sectional samples for transmission electron microscopy (TEM), scanning TEM, scanning electron microscopy and scanning ion microscopy. It is absolutely required for FIB cross-sectioning to prepare higher-quality samples in a shorter time without sacrificing the site specificity. The present paper clarifies the parameters that impose limitation on the following performances of the FIB cross-sectioning: milling rate, cross-sectioning at a right angle with respect to the sample surface, curtain structures formed on the cross sections, ion implantation and ion damage. All of these are discussed from the viewpoint of ion-sample interaction. Improvements for these performances achieved by diminishing their limiting origins or by correcting the resultants are described. Especially, the FIB scanning speed is significantly utilizable to improve the milling rate. A microsampling method, which allows the FIB incidence in a sidewards or upwards direction as well as downwards with respect to the microsample surface, is very effective to minimize the curtain structures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.