Abstract

We present a laboratory system designed for studying frost heave in fine-grained soil. The system consists of: a modified refrigerator, a frost heave test cell, a laser for measuring heave, a differential pressure transducer for measuring water intake, and platinum resistance temperature detectors for measuring pedestal temperatures. The frost heave cell allows for visual observation of the sample, and accommodates pretest sample consolidation, freezing tests using a variety of freezing methods, triaxial tests on frozen soil, and thaw consolidation tests. The modified refrigerator maintains the specified temperature ±0.5°C during the full length of the test. Test results indicate repeatability of frost heave ratios ξ to within ±7% , and average heave rates to within ±0.05 mm∕h . Results from frost heave tests conducted on five fine-grained soils indicate that: (1) a soil removed of its colloidal organic content becomes less frost susceptible; (2) the geomorphologic history of a “regional” soil is a critic...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.