Abstract

A ferrous gel, based on ferrous (Fe) sulphate and agarose, was used with a clinical magnetic resonance imaging (MRI) scanner to obtain relative dose distribution data from therapeutic photon and electron beams. The FeMRI gel was scanned using a new MRI acquisition protocol optimized for T1 measurements. Thorough comparisons with silicon semiconductor detector and ionization chamber measurements, as well as with Monte Carlo calculations, were performed in order to quantify the improvements obtained using FeMRI for dose estimations. Most of the relative doses measured with FeMRI were within 2% of the doses measured with other methods. The larger discrepancies (2-4%) found at shallow depths are discussed. The uncertainty in relative dose measurements using FeMRI was significantly improved compared with previously reported results (5-10%, one standard deviation, 1 SD), and is today between 1.6% and 3.3% (depending on dose level, 2 SD). This corresponds to an improvement in the minimum detectable dose (3 SD above background) from approximately 2 Gy to better than 0.6 Gy. The results obtained in this study emphasize the importance of obtaining basic FeMRI dose data before the method is extended to complicated treatment regimes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.