Abstract

In order to improve the detection efficiency and sensitivity for 236U measurement at China Institute of Atomic Energy (CIAE) accelerator mass spectrometry (AMS) system to meet the requirement of some applications, we present some new improvements of the measurement method for AMS measurement of 236U at CIAE AMS. The main features of the method include: (i) measurement optimization; (ii) improvement of the TOF time resolution ∼500ps. Based on these improvements, the sensitivity is ∼10−11 for 236U/238U in the present work, higher than before (236U/238U∼5 × 10−10).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.