Abstract

Standardisation of near-infrared spectrometric instruments using the piecewise direct standardisation method is studied. Two important steps of this method, namely the selection of the subset samples and the determination of the local optimal number of principal components, can yield artefacts in the transferred spectra, if they are not carefully performed. Three different algorithms are tested on two different data sets in order to study the subset selection and a method based on the prediction of an independent test sample is proposed in order to detect the artefacts due to bad local rank determinations. Improvements of the standardisation procedures were obtained, particularly when low numbers of standardisation samples were included in the subset.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.