The performance of medium-quality ion chromatographic equipment was improved by using pseudo-random sample intoduction correlation chromatography (CC). The importance of the digital preprocessing of the detector signal prior to decorrelation was demonstrated for realization of the multiplex advantage of the CC method. The detection limits obtained for single-column ion chromatography using conductivity were 0.1, 0.12, and 0.17 ppm for Cl −, NO 3 − and SO 4 2− ions, respectively. Using the same apparatus without CC the detection limits are in the range 1–3 ppm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.