Abstract

BackgroundIn stereotactic radiosurgery, sharp beam edges have clear advantages to spare normal tissues. In general, the dose gradient is a limiting factor in minimizing dose to nearby critical structures for clinical cases. Therefore the penumbral width should be diminished. MethodsA Varian Clinac 2100 linear accelerator equipped with in-house designed radiosurgical collimator was modeled using the EGSnrc/BEAMnrc Monte Carlo code and compared with the measurements. The 0.015 cm3 PinPoint chamber was used to measure the 6 MV photon beam characteristics and to validate Monte Carlo calculations. Additional to the standard (STD) linac, a flattening filter free (FFF) linac was simulated. Percent depth doses, beam profiles and output factors were calculated for small field sizes with diameter of 5, 10, 20 and 30mm with DOSXYZnrc. The mean energy and photon fluence at the water surface were calculated with BEAMDP for both FFF linac and STD linacs. ResultsThe penumbra width (80%-20%) was decreased by 0.5, 0.3, 0.2 and 0.2mm for field sizes of 5, 10, 20 and 30mm respectively when removing the FF. The fluence of photons at the surface increased up to 3.6 times and the mean energy decreased by a factor of 0.69 when removing the FF. The penumbra width (80%-20%) decreased by 17% when a 2 MeV monoenergetic electron pencil beam incident on the target is used instead of 6.2 MeV. ConclusionsIt was found that the penumbra of small field sizes is decreased by removing the FF. Likewise using low megavoltage photons reduced the beam penumbra maintaining adequate penetration and skin sparing.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call