Abstract

Studies with low-energy gamma-radiation show a dead layer of a thickness of about 1 mm at the surface of the intrinsic region of Ge(Li) detectors. Predominantly due to hole trapping in dead layer, low-energy gamma-spectroscopy below 100 keV with great-volume high-resolution Ge(Li) detectors is strongly degraded. A distinct improvement of the charge collection properties of the detector for low-energy gamma-quanta is achived, when the instrinsic region is simultaneously illuminated with infrared light in the energy range 0.69–0.82 eV. Consequently, utilizing this technique, it becomes possible to use great-volume Ge(Li) detectors for gamma-spectroscopy down to energies of about 30 keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.