Abstract

We have previously pointed out a defect of automatic endoscopic reprocessors, i. e. the contamination of the connecting part between the endoscope and the reprocessor. We evaluated a newly designed connector (MH-861; Olympus, Tokyo, Japan) with a self-cleaning and disinfection mechanism, which enabled cleaning and disinfection of both the connector itself and its interface with the suction and air/water valves during a reprocessing cycle, which was not previously possible. Ten upper gastrointestinal endscopes were examined in the study. Swabs were taken from the suction and air/water valves for microbiological culture before and after reprocessing by the washer-disinfector. The numbers of contaminated endoscopes before and after reprocessing with the new connector were compared. Before the procedure there were five contaminated endoscopes and none after the procedure. When the new connector was used, the difference in cleaning and disinfection of the connecting parts was significant (P = 0.0325). We conclude that the newly developed connector permits effective cleaning and disinfection by automatic reprocessors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.