Abstract

For high-energy-resolution photoelectron spectroscopy using synchrotron radiation, the energy resolution of a commercial compact photoelectron spectrometer (hemispherical concentric spectrometer) was improved by reducing the size of the entrance and detector slits and optimizing the operation conditions of the lens voltage. Under the optimized conditions, ray-tracing simulations show that severe spectral intensity decreases can be avoided. An energy resolution of 6.2 meV and a resolving power of 8100 at a kinetic energy of 50 eV were experimentally obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.