Abstract
BackgroundWe compared the cell counting accuracy of the conventional method and the improved method by using Neubauer counting chamber.MethodsIn the improved method, all the border cells were counted and then divided by two; while, in the conventional method, only border cells on the two boundaries (top and left) were counted.ResultsAbout 55.814% of the samples showed more accurate results by improved counting method, about 38.372% had more accurate results by conventional counting method, and about 5.814% were counted with similar counting error by both methods. The improved method had significantly smaller counting error than conventional method (P < .05). The distribution ratio of the border cells was an independent factor for counting accuracy (P < .05).ConclusionTogether, the improved counting method can reduce the counting error of the Neubauer counting chamber to some extent, assess the distributing uniformity of border cells, and help to eliminate the samples with large differences in distribution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.