A thin beryllium disk coated with a gold evaporated film of about 80 nm thickness was fabricated as an effective radiation window of bremsstrahlung X-rays. Its characteristics were examined in detail. The bremsstrahlung counting rate was proportional to the partial pressure of tritium above 0.1 Pa. The specific counting rate was evaluated as 70.8 cps/Pa. On the other hand, it decreased gradually with the total pressure above 2 kPa. This behavior was interpreted in terms of the self-absorption of β-particles in gas phase. It was revealed that the counting rate could be reproduced very well by a simple exponential function. From these results, it was concluded that the beryllium window coated with a thin gold evaporated film is a powerful device for in-situ measurements of tritium concentration in a wide range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.