Abstract

An automatic HPLC system for analyzing nitropolycyclic aromatic hydrocarbons (NPAHs, nitroarenes) in airborne particulates was previously described (Anal. Chim. Acta, 2001, 445, 20). Some problems with this system were that it generated a peak originating from an ascorbic acid solution that elutes at a retention time close to that of 1,6-dinitropyrene (DNP), and that it was able to analyze only 1,3-, 1,6-, 1,8-DNPs and 1-nitropyrene (I-NP). Here, we describe an improved system that effectively removes the interfering peak by introducing an ODS column just after the pump for the ascorbic acid solution, and which is capable of analyzing several additional compounds (2-, 4-NPs, 2-nitrofluorene. 6-nitrochrysene, 7-nitrobenz[a]anthracene, 3-nitroperylene and 6-nitrobenzo[a]pyrene etc.). The improved sensitivities were achieved by concentrating the compounds in a benzene-ethanol extract from airborne particulates, by increasing the loading time of the sample solution from 20 to 38 min, and by increasing the flow rate of an ascorbic acid solution from 1.3 to 1.8 mL/min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.