Abstract

The precision in measurement of trace level uranium isotopic ratio, i.e., 236U/238U or 234U/238U, on single Faraday detector with narrow dynamic range is very hard to achieve. this is mainly due to the narrow dynamic range of a single detector systems. A significant improvement in mass spectrometric determination of 236U/238U ratio has been achieved by employing an alternate method using a single Faraday detector of narrow dynamic range. The method makes use of the precise measurements of the 236U/234U ratio, 234U/235U ratio and 235U/238U ratio, which are used to calculate the 236U/238U ratio using the equation 236U/238U=236U/234U×234U/235U×235U/238U. Despite the fact that correlation of the data tends to increase the uncertainty in the result, our results show a significant improvement, i.e., more than 8 times better precision in measuring the 236U/238U ratio with this method (σ=3.98×10−08) as compared to direct measurement of 236U/238U (σ=3.104×10−07). The method widens the applicability of the single collector system with narrow dynamic range and it will potentially be helpful to improve the precision in the case of the static multi-collector system also. The objective of the present study was to compare the results of the same sample analyzed with the present alternate method and the direct method for precision.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.