Abstract

Currently most state-of-the-art setups for neutron radiography use scintillator screens and CCD cameras for imaging. However, in some situations it is not possible to use a CCD and alternatives must be considered. One such alternative is the well-established technique of track-etch neutron radiography, which has as main disadvantages requiring a long time for image recording and generating images with a relatively low contrast. In this work we address these negative issues and report significant improvements to recording and digitizing images using an improved setup consisting of an enriched 10B converter, a CR-39 solid state nuclear track detector and a flatbed scanner. The improved setup enables a significant reduction of the fluence required to obtain a neutron radiography image using this technique. Comparisons are made with imaging using two CCD models in the same beam line, so that the results can be extrapolated for other facilities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.