The total-body irradiation (TBI) technique at St Bartholomew's Hospital has been developed to improve dose homogeneity within the patient. Using a standard 6 MV linear accelerator in an orthodox-sized treatment room, the midpoint doses in head, neck, shoulders, mid-mediastinum, pelvis, knees and ankles are +/- 5% of that of the umbilicus in our current technique. This homogeneity has been achieved by a four-field technique, a reproducible patient set-up, careful use of a new bolus material and an additional beam-flattening filter mounted near the machine head. In addition, thermoluminescent dosimetric data collected at a test irradiation before TBI are used to influence field weightings and further improve dosimetry. This individualised and empirical TBI technique has dosimetric advantages over theoretical TBI dosimetric considerations in reducing dose gradients within the patient. These advantages are discussed.