Abstract

Fusion diagnostics that utilise high speed scintillators often need to capture a large area of light with a high degree of time accuracy. Microchannel plate (MCP) photomultiplier tubes (PMTs) are recognised as the leading device for capturing fast optical signals. However, when manufactured in their traditional proximity focused construction, the time response performance is reduced as the active area increases. This is due to two main factors: the capacitance of a large anode and the difficulty of obtaining small pore MCPs with a large area. Collaboration between Photek and AWE has produced prototype devices that combine the excellent time response of small area MCP-PMTs with a large active area by replacing the traditional proximity-gap front section with an electro-optically focused photocathode to MCP. We present results from both single and double MCP devices with a 40 mm diameter active area and show simulations for the 100 mm device being built this year.