Abstract
In high-field NMR microscopy rapid single-shot imaging methods, for example, echo planar imaging, cannot be used for determination of the apparent diffusion tensor (ADT) due to large magnetic susceptibility effects. We propose a pulse sequence in which a diffusion-weighted spin-echo is followed by multiple gradient-echoes with additional diffusion weighting. These additional echoes can be used to calculate the ADT and T 2 ∗ maps. We show here that this results in modest but consistent improvements in the accuracy of ADT determination within a given total data acquisition time. The method is tested on excised, chemically fixed rat spinal cords.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.