Abstract

At the present time, the relatively high atrial stimulation threshold and the low and often unstable P wave amplitude are the main concerns in atrial pacing. As a further contribution to the solution of this problem we used a screw-in atrial lead in 16 patients. During the implantation, atrial mapping was performed in order to establish the position with the best electrical performance. The acute threshold for voltage at a pulse duration of 1 ms averaged 0.49 +/- 0.18 Volt. The acute peak-to-peak P wave amplitude averaged 4.6 mV +/- 1.3. No complication occurred at the time of the implantation. Follow-up periods averaged 12 months. No dislocation or bleeding was observed during a follow-up period of up to 28 months. Using this lead together with intraoperative atrial mapping, it is possible to achieve an acute strength-duration relationship in the atrium at a level similar to that in the ventricle. The results of this study demonstrate that intraoperative mapping is an important stein in enhancing the safety margins and long-term performance of atrial pacing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.