Abstract

This study aimed to assess the impact of super-resolution deep learning reconstruction (SR-DLR) on coronary computed tomography angiography (CCTA) image quality and blooming artifacts from coronary artery stents in comparison to conventional methods, including hybrid iterative reconstruction (HIR) and deep learning-based reconstruction (DLR). A retrospective analysis included sixty-six CCTA patients from July to November 2022. Major coronary arteries were evaluated for image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). Stent sharpness was quantified using 10-90% edge rise slope (ERS) and 10-90% edge rise distance (ERD). Qualitative analysis employed a 5-point scoring system to assess overall image quality, image noise, vessel wall, and stent structure. SR-DLR demonstrated significantly lower image noise compared to HIR and DLR. SNR and CNR were notably higher in SR-DLR. Stent ERS was significantly improved in SR-DLR, with mean ERD values of 0.70 ± 0.20 mm for SR-DLR, 1.13 ± 0.28 mm for HIR, and 0.85 ± 0.26 mm for DLR. Qualitatively, SR-DLR scored higher in all categories. SR-DLR produces images with lower image noise, leading to improved overall image quality, compared with HIR and DLR. SR-DLR is a valuable image reconstruction algorithm for enhancing the spatial resolution and sharpness of coronary artery stents without being constrained by hardware limitations. The overall image quality was significantly higher in SR-DLR, resulting in sharper coronary artery stents compared to HIR and DLR.