RATIONALE Matrix-assisted laser desorption/ionization (MALDI) is one of the major techniques for mass spectrometry imaging (MSI) of biological systems along with secondary-ion mass spectrometry (SIMS) and desorption electrospray mass spectrometry (DESI). The inherent variability of MALDI-MSI signals within intact tissues is related to the heterogeneity of both the sample surface and the matrix crystallization. To circumvent some of these limitations of MALDI-MSI, we have developed improved matrices for lipid analysis based on structural modification of the commonly used matrix 2,5-dihydroxybenzoic acid (DHB). METHODS We have synthesized DHB containing -C6H13 and -C12H25 alkyl chains and applied these matrices to rat brain using a capillary sprayer. We utilized a Bruker Ultraflex II MALDI-TOF/TOF mass spectrometer to analyze lipid extracts and tissue sections, and examined these sections with polarized light microscopy and differential interference contrast microscopy. RESULTS O-alkylation of DHB yields matrices, which, when applied to brain sections, follow a trend of phase transition from crystals to an oily layer in the sequence DHB → DHB-C6H13 → DHB-C12H25. MALDI-MSI images acquired with DHB-C12H25 exhibited a considerably higher density of lipids than DHB. CONCLUSIONS Comparative experiments with DHB and DHB-C12H25 are presented, which indicate that the latter matrix affords higher lateral resolution than the former. Copyright © 2014 John Wiley & Sons, Ltd.