Abstract
In conventional spoiled gradient echo imaging utilizing quadratic radio frequency (RF) spoiling, nonideal signal intensities are often generated, particularly when repetition time is short and/or excitation flip angle (FA) becomes larger. This translates to significant errors in various quantitative applications based on T(1)-weighted image intensities. In this work, a novel spoiling scheme is proposed, based on random gradient moments and RF phases. This scheme results in a non-steady-state condition, but achieves ideal mean signal intensity. In order to suppress artifacts created by the inter-TR signal variations and at the same time attain the ideal signal intensity, radial data acquisition is utilized. The proposed method achieves ideal spoiling for a wide range of T(1), T(2), TR, and FAs. Phantom and in vivo experiments demonstrate improved performance for T(1) mapping and FA correction when compared with conventional RF spoiling methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.